Chapter 5: Maintaining Your Washer
Refer to Figure 3-7
in Chapter 3: Getting
Started for an
illustration showing
how to remove the
lower panel of your
washer.
54
Chapter 2: Prerequisites
To clean the water fill valve:
1. Turn off the water and electricity to the washer.
Unplug the Mobile modeld turn off the circuit breaker
that services the Undercounter or Freestanding model.
2. Remove the two screws at the bottom of the lower
panel of your washer.
3. Remove the panel by lifting slightly and pulling out.
4. Place a pan under the fill valve.
5. Remove the four screws from the valve plate and
separate the valve from the inlet connection.
6. Drain the water into the pan.
7. Remove the valve filter screen and clean it under
running water.
8. Replace the screen and reassemble the valve
connection. Be certain the O-ring gasket, identified in
Figure 5-1, is positioned in the groove in the fill valve
body.
9. Tighten all screwsd replace the lower panel, restore
electrical power to the washer, and turn on the water.
10. Check for leaks and adjust as needed. Replace the
lower panel.
Maintaining the Overflow
Dome
The overflow dome is located inside the washer in the
front left corner, as illustrated in Figure 5-2. The dome
may require periodic cleaning to remove residue from
inside the dome.