③ Set the range to display the spectrum diagram.
④ To specify the range with the time, click [Time] option and type the times.
⑤ Type the step to display the spectrum diagram in [Step] text box.
⑥ Type the channel number to display the spectrum diagram in [Ch to Display] text box.
⑦ Click【OK】button.
2-30
To specify the range with the cursors, click [Between Two Cursors] and click the
options of the cursor with which you specify the range.
・ Specify the range between 4th cycle or later from the beginning of the waveforms and the 4th cycle or
before the end the waveforms to display the spectrum diagram.
・ Type 0.4167 ms or over for the step value.
・ Divides the Range setting by the Step and the given value must be not more than fifteen.
・ The frequency spectrum diagram is displayed.
⎯ 142 ⎯