On Cleaning the Monitor

During normal use of the PDC the device may become dirty and should be regularly
cleaned.
The protection plate surface is especially treated to reduced reflection of light. When
solvents such as benzene or thinner, acid, alkaline or chemical cleaning cloth is used
for the protection plate surface, the performance of the monitor may be impaired or
the finish of the surface may be damaged. Take care with respect to the following
steps and suggested cleaning agents listed in the table C.1:
Steps:
1.
Prepare cleaning agent per manufacturer's instructions or hospital protocol.
2.
Wipe the POC with a clean cloth that has been moistened in the cleaning solu-
tion.
3.
Wipe thoroughly with a clean cloth.
Table C.1: Cleaning Agent list: Chemical Disinfectants which Have Been
Tested on the PDC-190S
No
1
2
3
4
5
6
7
8
9
10
!
Caution!
!
!
!
PDC-190S User Manual
Cleaning Agents
Cidex
Isopropyl alcohol
Green tinctured soap
Windex
Alcohol
Alcohol 70%
Chloride 1000PPM
Incidin plus
Incidin liquid
Mikrozid liquid
Do not immerse or rinse the PDC-190S or its peripherals. If you
accidentally spill liquid on the device, disconnect the unit from the
power source. Contact your Biomedical Department regarding the
continued safety of the unit before placing it back in operation.
Do not spray cleaning agents on the chassis.
Do not use disinfectants that contain phenol.
Do not autoclave or clean the PDC-190S or its peripherals with
strong aromatic, chlorinated, ketone, ether, or Esther solvents,
sharp tools or abrasives. Never immerse electrical connectors in
water or other liquids.