Calibration
Calibrating for Readback Accuracy
Calibrating for Readback Accuracy
The factory calibrates the offset and range of the output voltage and current monitor
circuits to within 1% for the default 0-5 Vdc scales. Recalibration may be necessary
when you change SW1 switch settings to select the 0-10 Vdc scale or when you
switch back to the 0-5 Vdc scale after previously calibrating for 0-10 Vdc operation.
Make calibration adjustments using multi-turn potentiometers, accessible through
holes in the cover of the power supply. Refer to
the pots.
Figure 5.3 Voltage Current and Monitor Calibration Locations
80
WARNING
Use a non-conducting, straight-bladed screwdriver to adjust the trim pots.
Figure 5.3
for the location of each of
Operating Manual for XHR Series Power Supply