Local Operation
Using Over Voltage Protection (OVP)
Resetting the
To reset the OVP circuit after it activates:
OVP Circuit
1. Reduce the power supply's output voltage setting to below the OVP set point.
2. Press the STANDBY switch IN. The S/D (Shutdown) LED on the front panel
3. Press the STANDBY switch again to return power to the load and resume normal
OR
1. Reduce the power supply's output voltage setting to below the OVP set point.
2. Turn the power supply OFF using the power switch, wait until the front panel
60
will turn on. The OVP LED turns off.
operation.
displays go out, then turn the supply back ON again.
Operating Manual for XHR Series Power Supply