Local Operation
Using Over Voltage Protection (OVP)
Resetting the
To reset the OVP circuit after it activates:
OVP Circuit
1. Reduce the power supply's output voltage setting to below the OVP set point.
2. Press the STANDBY switch IN. The red S/D LED on the front panel turns on.
3. Press the STANDBY switch again to return power to the load and resume
OR
1. Reduce the power supply's output voltage setting to below the OVP set point.
2. Turn the power supply OFF using the AC power switch, wait until the front
62
The OVP LED turns off.
normal operation.
panel displays go out, then turn the power supply back ON again.
Operating Manual for XFR 1.2kW Series Power Supply