To specify the probe skew

When measuring time intervals in the ns range, small differences in cable
length can affect the measurement. Use Skew to remove cable- delay errors
between any two channels.
1 Probe the same point with both probes.
2 Press one of the probes associated channel key.
3 In the Channel Menu, press Probe.
4 In the Channel Probe Menu, press Skew; then, select the desired skew
The skew setting is not affected by recalling the [Default Setup] or by
pressing [Auto Scale].

To calibrate a probe

See
InfiniiVision 7000B Series Oscilloscopes User's Guide
value.
Each analog channel can be adjusted ±100 ns in 10 ps increments for a
total of 200 ns difference.
"Calibrating Probes"
on page 57.
Displaying Waveforms
3
83