5
Measurements and Math Functions
When active, the Y1 and Y2 cursor values for the selected waveform
source are displayed in the softkey menu area.
The difference between Y1 and Y2 (Y) is displayed on the dedicated line
above the softkeys or in the display area when some menus are selected.

To make cursor measurements

If you intend to make cursor measurements on a trace that you recall from memory, be sure
N O T E
to recall both setup and trace. See
page 216.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Press the [Cursors] key. The key will illuminate, indicating that cursors
3 In the Cursors Menu, press Mode; the, select the desired mode:
178
are "on". (Press the key again when you want to turn cursors off.)
Manual — X, 1/X, and Y values are displayed. X is the
difference between the X1 and X2 cursors and Y is the difference
between the Y1 and Y2 cursors.
Track Waveform — As you move a marker horizontally, the vertical
amplitude of the waveform is tracked and measured. At the bottom
of the display, the time and voltage positions are shown for the
markers. The vertical (Y) and horizontal (X) differences between the
markers are shown as X and Y values.
"To recall waveform trace and/or oscilloscope setup"