For a 4- channel oscilloscope, connect BNC tees to the equal- length cables
as shown below. Then connect a BNC(f)- to- BNC(f) (barrel connector) to
the tee as shown below.
Figure 41
To Channel 2
3 Connect a BNC cable (40 inches maximum) from the TRIG OUT
4 Press the [Utility] key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear- panel CALIBRATION

To display the user calibration status

Pressing [Utility] > Service > User Cal Status displays the summary results of
the previous User Cal, and the status of probe calibrations for probes that
can be calibrated. Note that passive probes do not need to be calibrated,
but InfiniiMax probes can be calibrated. For more information about
calibrating probes see
400
User Calibration cable for 4-channel oscilloscope
To Channel 1
connector on the rear panel to the BNC barrel connector.
switch to PROTECTED.