CAN/LIN Triggering and Serial Decode
CAN Serial Decode
To set up the oscilloscope to capture CAN signals, see
Signals"
For CAN triggering set up see
N O T E
To set up CAN serial decode:
1 Press [Serial]; then, select the CAN serial decode mode.
2 If the decode line does not appear on the display, press the [Serial] key
3 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the
N O T E
oscilloscope is AutoTriggering. Press the [Mode/Coupling] key, then press the Mode
softkey to set the trigger mode from Auto to Normal.
You can use the Zoom horizontal sweep mode for easier navigation of the
decoded data.
266
on page 262.
to turn it on.
decode data.
"CAN Triggering"
on page 264.
InfiniiVision 7000B Series Oscilloscopes User's Guide
"Setup for CAN