4
Triggering
> qualifier time set softkey
• When the greater than (>) qualifier is selected, the Entry knob sets the
• When the within time range (><) qualifier is selected, the Entry knob
• When the out of time range (<>) qualifier is selected, the Entry knob
• When Timeout qualifier is selected, the Entry knob sets the timeout
When the duration trigger occurs
N O T E
The timer starts on the last edge that makes the pattern (logical AND) true. The trigger will
occur on the first edge that makes the pattern false if the time qualifier criteria of the
pattern has been met, except in Timeout mode. In Timeout, the trigger occurs when the
timeout value is reached while the pattern is true.
116
oscilloscope to trigger on a pattern duration greater than the time value
displayed on the softkey.
sets the lower time range value.
sets the upper time range value.
value.