To Perform Service Functions; User Calibration - Agilent Technologies MSO6012A User Manual

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

2
Front-Panel Controls

To perform service functions

User Calibration

102
• Press the Utility&Service to display the service menu.
Start
Start
User Cal
Self Test
The Service Menu lets you:
• Perform User Cal on the oscilloscope.
• View User Cal status.
• Perform instrument Self Test.
• View information about your oscilloscope model number,
code revision information, and User Cal status.
Perform user-calibration:
• Each year or after 2000 hours of operation.
• If the ambient temperature is >10° C from the calibration
temperature.
• If you want to maximize the measurement accuracy.
The amount of use, environmental conditions, and experience
with other instruments help determine if you need shorter User
Cal intervals.
User Cal performs an internal self-alignment routine to
optimize the signal path in the oscilloscope. The routine uses
internally generated signals to optimize circuits that affect
channel sensitivity, offset, and trigger parameters. Disconnect
all inputs and allow the oscilloscope to warm up before
performing this procedure.
About
User Cal
Oscilloscope
status
6000 Series Oscilloscope User's Guide
Return to
previous menu