MEASURING PHASE USING THE CURSORS

91
92
In this example you will use the cursors to measure the
phase difference between the waveforms on INPUT A and
B. Use the two waveforms from the dual input three-phase
measurement in the previous exercise. Cursors allow you
to make accurate digital measurements. It is advised that
you first press
to freeze the waveforms to be
measured.
Measure the two sine wave waveforms as explained
in the previous example.
Return to the main menu.
Set the test tool to Scope mode.
Freeze the display; HOLD appears in the upper right
display.
Activate the Cursor readings.
Go to page 2 within the More Scope window.
91/92/96/99/105
Choose ADD MAIN READING.
Open the list box.
: phase A>B
Close the window.
11
Position the two cursors on the waveforms; see
Figure 7-12.
Users Manual
.