SELECTING THE PROBE TYPE
To prevent measurement errors, ensure that your test tool
is adapted to your probe at power-on. Note that the
PM8918 probes, delivered with the test tool, have an
attenuation of 10:1.
For accurate measurement, periodic probe calibrations are
necessary. This is explained in Chapter 9, "User Mainte-
nance".
91/92/96/99/105
POWER-ON CONFIGURATIONS
There are two different ways to reset the ScopeMeter test
tool: the Master Reset, and the power-on/save memory.
Master Reset (Default Startup)
CAUTION
A Master Reset clears everything saved in memory.
Never perform a Master Reset unless you are prepared
to lose everything you saved in memory.
1. Turn the ScopeMeter test tool off.
2. Press and hold
.
3. Press and release
The test tool turns on, and you should hear a double
beep, indicating the Master Reset was successful. The
screen should display input: A PROBE 10:1 on top of
the waveform area.
4. Release
.
Perform a Master Reset only to make sure that your test
tool is in the initial settings condition. Master Reset sets the
test tool for using 10:1 probes.
See Chapter 6 for more information about Master Reset
conditions.