Advertisement
A "" statement identifies hazardous conditions and actions that could cause bodily harm or death.
A "" statement identifies conditions and actions that could damage the Meter or the equipment under test.
To avoid possible electric shock or personal injury, follow these guidelines:
Symbols
![]() | AC (Alternating Current) |
![]() | DC (Direct Current) |
Hazardous voltage | |
![]() | Battery (Low battery when shown on the display.) |
![]() | Do not dispose of this product as unsorted municipal waste. Contact Fluke or a qualified recycler for disposal. |
![]() | Fuse |
![]() | Double Insulated |
Important Information; Refer to manual | |
![]() | Earth ground |
![]() | AC and DC |
No. | Symbol | Meaning | Model |
![]() | ![]() | The Meter is in the VoltAlert™ non-contact voltage detect mode. | 117C |
![]() | ![]() | The Meter function is set to Continuity. | 115C & 117C |
![]() | ![]() | The Meter function is set to Diode Test | 115C & 117C |
![]() | ![]() | Input is a negative value. | 115C & 117C |
![]() | ![]() | | 115C & 117C |
![]() | ![]() | Display hold enabled. Display freezes present reading. | 115C & 117C |
![]() | ![]() MAX MIN AVG | MIN MAX AVG mode enabled. Maximum, minimum, average or present reading displayed | 115C & 117C |
![]() | (Red LED) | Voltage presence through the non-contact VoltAlert sensor | 117C |
![]() | LoZ | The Meter is measuring voltage or capacitiance with a low input impedance. | 115C & 117C |
![]() | nµF mVµA MkΩ kHz | Measurement units. | 115C & 117C |
![]() | DC AC | Direct current or alternating current | 115C & 117C |
![]() | ![]() | Battery low warning. | 115C & 117C |
![]() | 610000 mV | Indicates the Meter's range selection. | 115C & 117C |
![]() | (Bar graph) | Analog display. | 115C & 117C |
![]() | Auto Volts | The Meter is in the Auto Volts function. | 117C |
Auto | Auto ranging. The Meter selects the range for best resolution. | 115C & 117C | |
Manual | Manual ranging. User sets the Meter's range. | 115C & 117C | |
![]() | ![]() | Bar graph polarity | 115C & 117C |
![]() | ![]() | | 115C & 117C |
![]() | ![]() | | 115C & 117C |
No. | Description | Model |
![]() | Input terminal for measuring ac and dc current to 10 A. | 115C & 117C |
![]() | Common (return) terminal for all measurements. | 115C & 117C |
![]() | Input terminal for measuring voltage, continuity, resistance, capacitance, frequency and testing diodes. | 115C & 117C |
Error Messages | |
![]() | Battery must be replaced before the Meter will operate. |
![]() | Calibration required. Meter calibration is required before the Meter will operate. |
![]() | Internal error. The Meter must be repaired before it will operate. |
![]() | Internal error. The Meter must be repaired before it will operate. |
Switch Position | Measurement Function | Model |
![]() | Automatically selects ac or dc volts based on the sensed input with a low impedance input. | 117C |
Hz (button) | AC voltage from 0.06 mV to 600 V. Frequency from 5 Hz to 50 kHz. | 115C & 117C 115C & 117C |
![]() | DC voltage from.001 V to 600 V. | 115C & 117C |
![]() | AC voltage from 6.0 to 600 mV, dc-coupled. DC voltage from 0.1 to 600 mV. | 115C & 117C |
![]() | Ohms from 0.1 Ω to 40 MΩ. | 115C & 117C |
![]() | Continuity beeper turns on at <20 Ω and turns off at >250 Ω. | 115C & 117C |
![]() | Diode Test. Displays OL above 2.0 V. | 115C & 117C |
![]() | Farads from 1 nF to 9999 μF. | 115C & 117C |
Hz (button) | AC current from 0.1 to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off). Frequency from 45 Hz to 5 kHz. | 115C & 117C |
![]() | DC current from 0.001 A to 10 A (>10 to 20 A, 30 seconds on, 10 minutes off). >10.00 A display flashes. >20 A, OL is displayed. | 115C & 117C |
![]() | Non-contact sensing of ac voltage. | 117C |
Note: All ac functions and Auto-V LoZ are true-rms. AC voltage is ac-coupled. Auto-V LoZ, AC mV and AC amps are dc coupled. |
The Meter automatically enters "Sleep mode" and blanks the display if there is no function change, range change, or button press for 20 minutes. Pressing any button or turning the rotary switch awakens the Meter. To disable the Sleep mode, hold down the button while turning the Meter on. The Sleep mode is always disabled in the MIN MAX AVG mode.
The MIN MAX AVG recording mode captures the minimum and maximum input values (ignoring overloads), and calculates a running average of all readings. When a new high or low is detected, the Meter beeps.
To avoid electric shock, when Display HOLD is activated, be aware that the display will not change when you apply a different voltage.
In the Display HOLD mode, the Meter freezes the display.
Press to toggle the backlight on and off. The backlight automatically turns off after 40 seconds. To disable backlight auto-off, hold down
while turning the Meter on.
The Meter has both Manual and Autorange modes.
When you turn the Meter on, it defaults to Autorange and Auto is displayed.
To select a Power-Up Option, hold down the button indicated in the following table while turning the Meter on. Power-Up Options are canceled when you turn the Meter off and when sleep mode is activated.
Button | Power-Up Options |
![]() | Turns on all display segments. |
![]() | Disables beeper. ![]() |
![]() | Enables low impedance capacitance measurements. ![]() |
![]() | Disables automatic power-down ("Sleep mode"). ![]() |
![]() | Disables auto backlight off. ![]() |
The figures on the following pages show how to make basic measurements.
When connecting the test leads to the circuit or device, connect the common (COM) test lead before connecting the live lead; when removing the test leads, remove the live lead before removing the common test lead.
To avoid electric shock, injury, or damage to the Meter, disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
Note
The continuity function works best as a fast, convenient method to check for opens and shorts. For maximum accuracy in making resistance measurements, use the Meter's resistance (Ω) function.
With the function switch in the position, the Meter automatically selects a dc or ac voltage measurement based on the input applied between the V and COM jacks.
This function also sets the Meter's input impedance to approximately 3 kΩ to reduce the possibility of false readings due to ghost voltages.
With the function switch in the position, the Meter measures ac plus dc millivolts. Press
to switch the Meter to dc millivolts.
To avoid personal injury or damage to the Meter:
Turn circuit power off, break the circuit, insert the Meter in series with the circuit and then turn circuit power on.
The millivolt and voltage function of the Meter can be used with an optional mV/A output Current Probe to measure currents that exceed the rating of the Meter. Make sure the Meter has the correct function selected, AC or DC, for your current probe. Refer to a Fluke catalog or contact your local Fluke representative for compatible current clamps.
To avoid electrical shock, disregard the bar graph for frequencies >1 kHz. If the frequency of the measured signal is >1 kHz, the bar graph and are unspecified.
The Meter measures the frequency of a signal by counting the number of times the signal crosses a trigger level each second. The trigger level is 0 V, 0 A for all ranges.
Press to turn the frequency measurement function on and off. Frequency works with ac functions only.
In frequency, the bar graph and range annunciator indicate the ac voltage or current present.
Select progressively lower ranges using manual ranging for a stable reading.
To detect the presence of ac voltage, place the top of the Meter close to a conductor. The Meter gives an audible as well as visual indication when voltage is detected. There are two sensitivity settings. The "" setting can be used on flush mounted wall sockets, power strips, flush mounted industrial outlets and various power cords. The "
" setting allows for ac voltage detection on other styles of recessed power connectors or sockets where the actual ac voltage is recessed within the connector itself. The VoltAlert detector works in bare wire applications with voltages as low as 24 V in the "
" setting.
If there is no indication, voltage could still be present. Do not rely on the VoltAlert detector with shielded wire. Operation may be effected by differences in socket design, insulation thickness and type.
For making capacitance measurements on cables with ghost voltage, hold while turning on the Meter to switch the Meter into LoZ, (low input impedance) Capacitance mode. In this mode, capacitance measurements will have a lower accuracy and lower dynamic range. This setting is not saved when the Meter is turned off or goes into sleep mode.
The bar graph is like the needle on an analog meter. It has an overload indicator () to the right and a polarity indicator (
) to the left.
Because the bar graph is much faster than the digital display, the bar graph is useful for making peak and null adjustments.
The bar graph is disabled when measuring capacitance. In frequency, the bar graph and range annunciator indicates the underlying voltage or current up to 1 kHz.
The number of segments indicates the measured value and is relative to the full-scale value of the selected range.
In the 60 V range, for example (see below), the major divisions on the scale represent 0, 15, 30, 45, and 60 V. An input of −30 V turns on the negative sign and the segments up to the middle of the scale.
Test fuse as shown below.
Maintenance of the Meter consists of battery and fuse replacement, as well as case cleaning.
To avoid shock, injury, or damage to the Meter:
To remove the battery door for battery replacement:
To open the case for fuse replacement:
Wipe the case with a damp cloth and mild detergent. Do not use abrasives, isopropyl alcohol, or solvents to clean the case or lens/window. Dirt or moisture in the terminals can affect readings.
Accuracy is specified for 1 year after calibration, at operating temperatures of 18°C to 28°C, with relative humidity at 0% to 90%.
Extended specifications are available at www.Fluke.com.
Maximum voltage between any terminal and earth ground: 600 V
Surge Protection: 6 kV peak per IEC 61010-1 600V CAT III, Pollution Degree 2
Fuse for A input: 11 A, 1000 V 17 kA FAST Fuse (Fluke PN 803293)
Display: Digital: 6,000 counts, updates 4/sec
Bar Graph: 33 segments, updates 32/sec
Temperature: Operating: -10°C to + 50°C
Storage: -40°C to + 60°C
Temperature Coefficient: 0.1 x (specified accuracy)/°C (<18°C or >28°C)
Operating Altitude: 2,000 meters
Battery: 9 Volt Alkaline, NEDA 1604A / IEC 6LR61
Battery Life: Alkaline: 400 hours typical, without backlight
Safety Compliances: Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 No 61010-1-04, UL 6101B (2003) and IEC/EN 61010-1 2nd Edition for measurement Category III, 600 V, Pollution Degree 2, EMC EN61326-1
Certifications: UL, P, CSA, TÜV, (N10140), VDE
IP Rating (dust and water protection): IP42
Table 1. Accuracy Specifications
Function | Range | Resolution | Accuracy ± ([% of Reading] + [Counts]) | Model | |
DC millivolts | 600.0 mV | 0.1 mV | 0.5% + 2 | 115C, 117C | |
DC Volts | 6.000 V 60.00 V 600.0 V | 0.001 V 0.01 V 0.1 V | 0.5% + 2 | 115C, 117C | |
DC, 45 to 500 Hz | 500 Hz to 1 kHz | ||||
Auto-V LoZ[1] True-rms | 600.0 V | 0.1 V | 2.0% + 3 | 4.0% + 3 | 117C |
45 to 500 Hz | 500 Hz to 1 kHz | ||||
AC millivolts[1] Truerms | 600.0 mV | 0.1 mV | 1.0% + 3 | 2.0% + 3 | 115C, 117C |
AC Volts[1] True-rms | 6.000 V 60.00 V 600.0 V | 0.001 V 0.01 V 0.1 V | 1.0% + 3 | 2.0% + 3 | 115C, 117C |
Continuity | 600 Ω | 1 Ω | Beeper on < 20 Ω, off > 250 Ω; detects opens or shorts of 500 μs or longer. | 115C, 117C | |
Ohms | 600.0 Ω 6.000 kΩ 60.00 kΩ 600.0 kΩ 6.000 MΩ 40.00 MΩ | 0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ | 0.9% + 2 0.9% + 1 0.9% + 1 0.9% + 1 0.9% + 1 5% + 2 | 115C, 117C | |
Diode test | 2.000 V | 0.001 V | 0.9% + 2 | 115C, 117C | |
Capacitance | 1000 nF 10.00 μF 100.0 μF 9999 μF | 1 nF 0.01 μF 0.1 μF 1 μF | 1.9% + 2 1.9% + 2 1.9% + 2 100 μF - 1000 μF: 1.9% + 2 > 1000 μF: 5% + 20 | 115C, 117C | |
Lo-Z Capacitance (Power-up option) | 1 nF to 500 μF | 10% + 2 typical | 115C, 117C | ||
AC Amps True-rms[1] (45 Hz to 500 Hz) | 6.000 A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. | 0.001 A 0.01 A | 1.5% + 3 | 115, 117 | |
DC Amps | 6.000 A 10.00 A[3] 20 A for 30 seconds max., 10 minutes rest min. | 0.001 A 0.01 A | 1.0% + 3 | 115, 117 | |
Hz (V or A input)[2] | 99.99 Hz 999.9 Hz 9.999 kHz 50.00 kHz | 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz | 0.1% + 2 | 115, 117 |
Notes:
[1] All ac ranges except Auto-V LoZ are specified from 1% to 100% of range. Auto-V LoZ is specified from 0.0 V.
Because inputs below 1% of range are not specified, it is normal for this and other true-rms meters to display nonzero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of ≤3. AC volts is ac-coupled. AutoV LoZ, AC mV, and AC amps are dc-coupled.
[2] AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC Amps Hz is dc-coupled and specified from 45 Hz to 5 kHz.
[3] >10 A unspecified.
© 2006, 2007 Fluke Corporation, All rights reserved. Printed in China
All product names are trademarks of their respective companies.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the 115C and is the answer not in the manual?
Questions and answers