Fluke 91 Series II User Manual page 111

Scopemeter series ii
Table of Contents

Advertisement

6 - 18
2. Choose the type of TEST WAVEFORM: the test tool
outputs a slow voltage sweep (-2V to +2V) or current
sweep (0 to +3mA).
3. Choose the TEST PROBE COUPLING: AC or DC
coupling.
Connect the component to be tested between the red
(GENERATOR OUT) and black (COMmon) 4-mm banana
jacks. Measure the component on INPUT A. It's character-
istics are shown on the special measure display.
Auto Set cannot be used with the Component measure-
ment mode.
The given display shows a diode test with a slow voltage
sweep (-2 to +2V).
91/92/96/99/105
Users Manual