Calibration Parameters - Cardinal 225 Installation And Technical Manual

Weight indicator
Table of Contents

Advertisement

CALIBRATION AND SETUP, CONT.
Single-Point for Zero Only (Only Zero) Calibration
CAL 1= (FIRST CALIBRATION WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed, press the 1 and 0 keys and then the
ENTER key. The display will change to show the settings for capacity, interval, decimal
point position and CAL 1=0.0. This is the first of two calibration steps.
1
. Insure the scale is empty and then press the ENTER key.
2. The display will flash CALIBRA
show CAL 2=.
CAL 2= (SECOND CALIBRATION WEIGHT)
The display will show
1. Press the ZERO key.
2. The display will change to show the SCALE 1 CALIBRATION MENU.

Calibration Parameters

The calibration parameters (Swt, Sct and Zct) represent the test load (weight) and the
analog to digital readings at zero and test load. These values are established when the
scale is calibrated. To return the indicator to its present calib
test we
ights) enter the calibration paramete
1. Swt=XXXXXXX (SPAN WEIGHT)
With the SCALE 1 CALIBRATION MENU displayed the current setting for the Swt=
parameter will be shown. Note that XXXXXXX is the current value for the calibrated te
load Span weight. If
parameter.
Otherwise, press the 1 key, the ENTER key and then using the numeric keys, enter th
value for the
the new setting and return to the SCALE 1 CALIBRATION MENU. Allowable values
1 through 9,999,999.
2. Sct=XXXXXXX (SPAN COUNT)
With the SCALE 1 CALIBRATION MENU displayed the current setting for the Sct=
parameter will be shown. Note that XXXXXXX is the current value for the analog to
digital reading at test load Span count. If the setting displayed is accepta
the next setup parameter.
Otherwise, press the 2 key, the ENTER key and then using the numeric keys, enter the
value for the test load Span count and then press the ENTER key to save the new
setting and return to the SCALE 1 CALIBRATION MENU. Allowable values are 1
through 9,999,999. Note that the capacity can not exceed 7 digits or 9,999,999.
8200-M538-O1 225 Installation
TING... for a few seconds and then change to
CAL 2=XXXXXXX. This is the second of two calibration steps.
the setting displayed is acceptable, proceed to the next setup
calibrated test load Span weight and then press the ENTER key to save
Note that the capacity
ration settings (without using
rs values.
can not exceed 7 digits or 9,999,999.
st
e
are
ble, proceed to
Page 69