Filter Setting Recommendations - Cardinal 225 Installation And Technical Manual

Weight indicator
Table of Contents

Advertisement

CALIBRATION AND SETUP, CONT.
N
on Critical
Sample Rate
rate is not critical, as in static weighing, set FILTER= to:
If the sample
FILT
ER=0 (NO Filtering)
FILT
ER=1 (FMAX=6, FBRK=12, SAMPLES=2/Sec) or
FILT
ER=2 (F
Critical Sample Rate
With a critical sample rate, as in a filling operation, use Custom Filtering (FILTER=3).
1. SAMPLES= S
Set the sample rate as close as possible to produce a display graduation change for
every graduation of material added to the scale.
Material Flow Rate
U
(lbs/second)
Interval
EXAMPLE:
2. FBRK
=
FILTER BRE
Turn the filtering off by setting FILTER=0
used and, by observation, determine the number of grads of instability that needs to be
f ered
ilt
out. Set the break range (FBRK=) to that value.
Weight Change
U
Interval
EXAMPLE: 20,000 x 10lb capacity scale with 800lb variation in the weight display.
800
U
U
10
3. F
MAX= MAXIMUM FILTER LEVEL (0 to 255) determination: Set to desired results.
f stability is unacceptable with any setting of FMAX=, reduce the sample r
4. I
increas
e t
he br
8200-M538-O1 225 Installation

Filter Setting Recommendations

MAX=6, FBRK=8, SAMPLES=1/Sec).
AMPLE RATE (1 to 200 samples/second) determination:
U
100lb
s/sec
U
U
10lbs
AK RANGE (1 to 64 graduations) determination:
= FBRK
U
= FBRK=80
eak range, FBRK= setting
= SAMPLES
= 10s/s = SAMPLES
. Operate the system
for increased filtering.
as it will be normally
ate and/or
Page 59