7.2

Shaking speed

The maximum rotation speed allowable depends on the position and the loading.

Single unit (universal tray with clamps)

7.2.1
Load with clamps
25 x 300 mL
16 x 500 mL
9 x 1000 mL
4 x 2000 mL
2 x 5000 mL
Load with «Sticky Stuff»
25 x 300 mL
16 x 500 mL
9 x 1000 mL
4 x 2000 mL
2 x 5000 mL
7.2.1

Stacked units

Upper deck
Load with clamps
Safe
Critical
Load with «Sticky Stuff»)
25 x 300 mL
16 x 500 mL
9 x 1000 mL
4 x 2000 mL
2 x 5000 mL
Lower deck
Same as single unit.
The following standard values must be observed:
If vibration occurs, reduce the speed until the equipment runs quietly. Alternatively, raise or lower
the loading until the equipment runs quietly.
The normal operating range lies between 200 up to 400 rpm. To reach higher rotational speeds (up
to 500 rpm) is perfectly possible; however the customer must communicate this wish to an INFORS
supplier beforehand so we can advise accordingly.
(Operating-Manual_Ecotron _e)
Safe
300 rpm
300 rpm
250 rpm
200 rpm
200 rpm
Safe
200 rpm
200 rpm
250 rpm
200 rpm
200 rpm
200 rpm
240 rpm
Safe
200 rpm
200 rpm
200 rpm
200 rpm
200 rpm
Critical
400 rpm
400 rpm
380 rpm
280 rpm
260 rpm
Critical
-
-
-
250 rpm
250 rpm
Critical
-
-
240 rpm
240 rpm
240 rpm
page 24 of 53