7.6.3
Settings
Event List
Info
27.05.2024
The event list shows the last 18 events including their timestamps
(date and time). The latest event is always at the top.
The following events are displayed in the event list:
Activating/deactivating parameters (manually on the device,
■
remotely or by the timer)
Changing of setpoints
■
Opening the door (incl. duration)
■
Alarms and error messages (alarms and error messages are addi-
■
tionally marked with a red exclamation mark symbol)
The event list consists of a maximum of three pages with six entries
each. Use the lower Plus or Minus key to scroll through the different
pages of the event list.
The event list can only be displayed locally on the device and cannot
be exported.
The event list does not correspond to an audit trail in
the sense of FDA 21 CFR Part 11. The data stored in the
event list can and may therefore not be used in the
regulated environment.
To record and store the data under GMP conditions,
you need eve
trail function.
Display of device and status information. The following information is
displayed:
Mainboard Firmware : Shows the version of the controller firmware
■
that is currently installed.
Display Firmware : Shows the version of the display firmware that is
■
currently installed.
Operating hours shaker : Shows the operating hours of the shaker.
■
Operating hours fans : Shows the operating hours of the fans.
■
Number of door openings : Shows how often the door of the
■
device has been opened.
Number of pulses humidifier pump : Shows how many switching
■
cycles (pulses) the humidity control pump (HHC) has run through.
This item is only displayed if the humidity control option (HHC) is
installed.