3.3
Openings
3.3.1
Discharge Outlet
3.3.2
Air Vents
27.05.2024
For draining spilled liquids, cleaning agents or accumulated condensa-
tion water, there is a discharge outlet at the bottom, almost in the
middle of the left side of the casing.
The opening is sealed with a yellow plug on delivery. A hose nozzle
(¼ inch) for connecting a hose (Ø 10 mm) is provided.
In case of large fill volumes, we recommend installing
the discharge hose to avoid the bearings coming into
contact with liquid if a flask breaks.
If safety requirements must be met, e.g. when
working with genetically modified organisms, the dis-
charge hose must lead to a suitable, sealed recep-
tacle. One example of such a receptacle would be an
empty chemical container sealed with foil.
The device has several air vents. When setting up the device, you must
ensure that these air vents remain unobstructed (keeping a minimum
distance of 80 mm to walls or objects on either side) in order to
ensure that the exit gas can be extracted without obstruction and
none of the components overheat.
Multitron - Operating manual
Setup and Function
Page 33 of 140