Advertisement

4.8
Cable Pass-through
4.8.1
Setup and Function
1
2
27.05.2024
A pass-through for cables and hoses can be installed on the right side
of the housing. This can be used to guide additional sensors or gassing
into the incubation chamber.
The pass-through reduces loss of heat and/or humidity and, if appli-
cable, gas consumption if cables or hoses have to be passed into the
incubation chamber.
The cable pass-through consists of a dividable cable-pass-through
bar (1) into which slotted sealing elements (cable sleeves) (2) can be
inserted. The cables are affixed and strain-relieved using cable sleeves.
Openings that are not required are sealed using a dummy sleeve.
The cable pass-through is screwed onto the inside of the housing. A
suitable screwdriver (SW 4) is included to dismantle the removable
part of the cable pass-through (3).
3
Multitron - Operating manual
Options
Page 49 of 140