Unsuitable Installation Sites
6.2
Requirements for the Mains Connection
27.05.2024
Installation and Commissioning
In order to achieve the desired climate in the incubation chamber
and to be able to control it as precisely as possible, the following
installation locations must be avoided at all costs:
Poorly ventilated recesses or areas in the exhaust heat air stream
of air conditioners or other sources of heat
Directly underneath, next to or in the air stream of the air condi-
tioners or other devices that generate strong air flows (e.g. circula-
tion fans in clean rooms)
Directly next to device with strong heat radiation, such as heaters
or autoclaves, or in their exhaust heat stream
Directly next to refrigeration devices, such as deep freezers or
refrigerated centrifuges, or in their exhaust heat air stream
To be able to achieve the desired temperature in
the incubation chamber, the ambient temperature
directly at the device is decisive. This can be consider-
ably higher than the temperature at other positions
in the room due to the waste heat of the unit or
due to other devices with strong heat radiation in the
immediate vicinity!
To avoid dangers due to electrical current, the in-house mains connec-
tion must meet the following requirements:
Constant power supply
Recommendation: In the building, secure the power supply using a
fault current protection switch (RCD – Residual Current Device).
In addition to that, ensure the following:
The voltage values of the device match those of the local mains
voltage. Note the information on the identification plate.
Use the power cable provided. If the power cable becomes defec-
tive, replace it with a power cable of the same type.
The mains connection can be accessed at all times.
Refer to the technical data for the electric connection values.