Step 22: Follow Manual Directions to Install All Software
Installation of the CTech ViPER ANLYTX Software should only be done by a
trained Support Specialist or under the guidance of one.
While logged into the computer with administrator privileges, install the ViPER
ANLYTX Software.
Step 23: Power on Cary 60
With the USB cable disconnected from the computer, press the Power button
on the front of the Cary 60. Power up the Cary 60 and allow it to fully initialize.
Step 24: Allow Cary 60 to Complete Start-up Initialization
When the Cary 60 is powered on, the hardware will initialize and calibrate, and
the Power button will light up and flash an orange color. After pressing the
Power button to turn on the Cary 60, wait for it to turn green. A steady green
light means the Cary 60 has been initiated successfully.
Note: If the System fails to initialize and calibrate successfully, the Power
button will turn red. If it fails to start-up successfully, turn it off and turn
it on again.