CTech™ SoloVPE® PLUS System
4. Load the filled sample vessel into the sample vessel holder and insert a clean, new Fibrette Optical Component into the
Quick Set Coupler. Slide the SoloVPE cover down to block stray light from reaching the detector.
5. Click OK. Wait for the System Suitability Test to finish.
6. Test results will be added to the System Suitability Log. If the test was initiated through the Quick Slope or AAV app, the test
result will be displayed in the report.
DOC0362 eRev. 1.0 [Publication Date]
Figure 24. Reference Standard Information Fields
User Guide
34