2-point adjustment

Display current values of the 2 points as any adjustment values,
respectively.
Display the measurement value with span-adjusted to any display value.
For 4-20 mA output ("6 4-20 mA Output Setting" P.14) and alarm output
("Cleaning output setting" P.17), the adjusted values are applied.
Electromotive force cannot be adjusted.
If offset, span, or 2-point adjustment has already been performed,
Caution
be sure to initialize the adjustment before performing the 2-point
adjustment.
Measurement item
(display range)
[Display]
Conductivity (0.00 to
10.00 mS/cm) [EC]
Conductivity (0.00 to
10.00) [EC]
Conductivity (0.000
to 1.000 S/m) [EC]
Conductivity (0.000
to 1.000) [EC]
Conductivity (0.0 to
30.0 g/L) [EC]
[Point 1]
Display value
after 2-point
adjustment
Adjusted
value
Current display
value
Setting range (display value/
adjustment value)
Point 1: 0.00 to (2nd point value -1.00)
Point 2: (1st point value +1.00) to 10.00
Point 1: 0.00 to (2nd point value -1.00)
Point 2: (1st point value +1.00) to 10.00
Point 1: 0.00 to (2nd point value -0.100)
Point 2: (1st point value +0.100) to 1.000
Point 1: 0.000 to (2nd point value -0.100)
Point 2: (1st point value +0.100) to 1.000
Point 1: 0.0 to (2nd point value -1.0)
Point 2: (1st point value +1.0) to 30.0
[Point 2]
Adjusted value
Current display
value
Measurement value
Factory
default
setting
Point 1: 0.00
Point 2: 10.00
Point 1: 0.00
Point 2: 10.00
Point 1: 0.00 0
Point 2: 1.000
Point 1: 0.00 0
Point 2: 1.000
Point 1: 0.0
Point 2: 30.0
23