Response Time - Optex SE-Cond-D Instruction Manual

Conductivity sensor
Table of Contents

Advertisement

Measurement
item
Main
Conductivity
display

Response time

Set a signal output response time.
A measurement value is calculated using moving average of the time period
set by the signal output response time. For example, if the sensor measures
values every 5 seconds and the signal output response time is set to 50
seconds, which is 5 seconds × 10 times, the moving average of 10 times
measurement is calculated and used as the measurement value.
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "6.1 Sensor Setting" and "Response Time".
Response
5 to 600 seconds (by 5 seconds)
Time
Display item
Item (display range)
EC (0.00 to 10.00 mS/cm) mS/cm EC (0.00 to
EC (0.00 to 10.00)
EC (0.000 to 1.000 S/m)
EC (0.000 to 1.000)
EC (0.0 to 30.0g/L)
Setting range
Factory default
setting
Unit
10.00 mS/cm)
-
mS/cm
S/m
-
g/L
Factory default
setting
5 seconds