2.3

How to install the STP pump

2
Install the STP pump to the vacuum equipment as shown in Figure 14.
Use a lifter or the like when installing the STP pump to the vacuum equipment. Failure to do so
might hurt your back or cause injuries due to occurrence of an accident such as fall.
A supportive device such as lifter to installing the STP pump should withstand the load of five times
or more the weight of the STP pump.
MT-00E-001-M
Page 20
STP Series Turbomolecular Pump
An appropriate enclosure or a barrier which cannot be removed without using a tool
should be provided to prevent an operator from accessing the connection cables
between the STP pump and its connectors provided.
When installing the STP pump and connecting/disconnecting cables, always power
off the primary power (switch the MAIN POWER to "OFF") and isolate
(Lockout/Tagout) the electrical energy source, water and gas, and other energy
sources on the vacuum equipment. Failure to do so may result in the inadvertent
rotation of the STP pump which may result in an accident, an electric shock or
Moreover, an accident caused by water leaks or gas leak may occur.
DO NOT remove the connecter while the pump is rotating. The voltage might be
output to connector according to the rotational speed, and it causes the electric
WARNING
WARNING
damage to equipment.
WARNING
shock or the failure.