CRYO-U
Cryopump Instruction Manual
®
Cryopump
5. Reduce the pressure in the refrigerator unit and the flexible hoses to a level of 0.3MPaG by
opening the valve(V1) of the charging adaptor.
NOTE:
As refrigeration lowers the temperature of helium gas in the refrigerator unit, the density of
the gas increases. After shutdown of the refrigerator, the temperature of the helium gas in
the refrigerator returns to room temperature from cryostatic temperature gradually, and the
inner pressure rises at the same time. The unit has a pressure relief valve which opens at
the inner pressure of 1.9MpaG and more. If once the pressure relief valve opens, it could
cause a leakage that would depend on the surrounding environment such as dust.
Therefore, ensure to reduce the inner pressure right after shutdown of the refrigerator in
order not to work the pressure relief valve.
Cryopump
Coldhead Cable
Disconnect here
Flexible Hose
SUPPLY
RETURN
Figure 8-2 Disconnecting Flexible Hoses
Flexible Hose
XXXXXXXX
******
Coldhead Cable
Figure 8-3 Connecting Charging Adopter
ULVAC CRYOGENICS INCORPORATED
INPUT POWER
COLD HEAD POWER
REMOTE
NO.1
NO.2
NO.3
RESPONSE
Power Cable
RETURN GAS
RETURN GAS
SUPPLY GAS
WATER IN
WATER OUT
OUT
Charging Adopter
V2
SUPPLY
V1
RETURN
V3
IN
Cooling Water Line
Release He gas
Release He gas
8-9