Using The Articulator Plate; Using Clamping Jaws - Dentsply Sirona inEos X5 Operating Instructions Manual

Hide thumbs Also See for inEos X5:
Table of Contents

Advertisement

12 Accessories

12.1 Using the articulator plate

62
12.1
Using the articulator plate
1. Use the adjusting screw to set the height of the articulator plate in
such a way that the model lies horizontally in the articulator on the
articulator plate.
2. Place the articulator plate in the working area of the inEos X5 in
such a way that the buccal scan can be taken.
12.2

Using clamping jaws

Accessories for the inEos X5 model holder for mounting smaller
models, quadrant models and individual stumps in the model holder.
The clamping jaws are available as STL files on the Dentsply Sirona
homepage in the inLab download area. If required, they can be
downloaded at the following address and then, for example, produced
additively or by milling.
https://www.dentsplysirona.com/de-de/entdecken/lab/cad-cam-fuer-das-
labor/downloads.html
Dentsply Sirona
Operating Instructions inEos X5
64 17 096 D3586
D3586.201.01.16.02    2023-12