Automatic Jaw Scan; Preparing The Optical Impression - Dentsply Sirona inEos X5 Operating Instructions Manual

Hide thumbs Also See for inEos X5:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions inEos X5
"Reduced Capturing" :
"Complete Capturing" :
64 17 096 D3586
D3586.201.01.16.02    2023-12
9.2

Automatic jaw scan

In "Capture Jaw" mode, you capture the model situation fully
automatically.
● For all tasks.
● Especially for large, complex tasks or tasks with especially high
accuracy requirements.
9.2.1

Preparing the optical impression

In the automatic scan, you can choose between the "Reduced
Capturing" and "Complete Capturing" options.
The entire jaw is scanned based on a standard sequence with a
lower level of detail.
Tip: This mode is especially suitable for scanning an unprepared op-
posite jaw. It provides all of the information required for the following
steps. Hard-to-scan gaps are not automatically closed.
The area scanning option is used to define the areas in which the
scan is automatically performed with a high level of detail. The areas
of the preparations are usually the ones involved here.
All of the gaps in the digital model are completely closed in the
scanned areas via the intelligent gap closing technique.
All other areas of the model are scanned with a lower degree of detail
using a standard sequence.
Model holding plates
Tip: If you are using partial jaw models (e.g. quadrant models), position
these on the outer edge of the model holding plate.
ü You have selected the desired object (lower jaw, upper jaw, etc.).
ü The articulating arm is moved to the loading position.
1. Fasten the model to the supplied model holding plate using the Blue
Tack filling material. The labial side must face the straight edge of
the plate.
or
➢ Alternatively, you can use the parallel vise provided. In this case,
the labial side faces the fastening screw.
9 Performing 3D acquisition
9.2 Automatic jaw scan
41