Mount The Power Supply; Interconnect Between Power Supply And Actuator - Branson 2000 Series Installation Manual

Hide thumbs Also See for 2000 Series:
Table of Contents

Advertisement

2000 Series
Installation Guide

Mount the Power Supply

The power supply is designed to be placed on a workbench (rubber feet on bottom) within cable-
length limits of the actuator, or it may be rack-mounted in a standard 19-inch Rack (using an
optional rack mount handle kit). It has two rear-mounted fans which draw cooling air from rear to
front, which must be free from obstruction. Do not place the power supply on the floor or in other
locations that will allow dust, dirt or contaminants to be drawn into the power supply.
The controls on the front of the power supply must be accessible and readable for setup changes.
All electrical connections are made to the rear of the power supply, which should be positioned in
your workspace with adequate clearance (approximately 4 inches or more on either side, and 6
inches to the rear) for cable access and ventilation. Do not place anything on top of the power sup-
ply case.
In the event the system is to be installed in a high dust environment, the use of a fan filter kit (2000
t, ea, d, and f: 101-063-614, 2000b/bdc, 101-063-696) is required.
See Figure 5 for a dimensional drawing of the 2000 Power Supply.

Interconnect between Power Supply and Actuator

Figure 20 Connections on Rear of Power Supply
Rear View of Power Supply,
showing connections and fans
100-214-226 Rev. 10
Fans
User I/O Connector
Actuator Interface Connector
Serial Port (RS-232) Connector for Host Computer
Parallel Printer Connector
(ultrasonic energy out)
DIP Switch for User I/O
Factory Air
RFConnector
Model Data Tag
Power Cord
27