OdaLog® L2 – Gas Logger User Manual
At low voltage, the battery symbol will be displayed as "empty", i.e., without any segments visible, as
shown in stage 3 above. Then, as the battery capacity decreases to a flat (non-operational, stage 4) state,
the OdaLog® instrument will stop the active logging session and switch the instrument off, continuously
displaying the word "FLAT" on the display with the "Battery Warning" symbol flashing.
Note:
The instrument cannot be turned back on until the battery has been replaced, please contact
Thermo Fisher/ Authorised Service Center.

6.12. Key Lock (Default = Disabled)

Key Lock (when enabled in the OdaStat-G™ software): this feature allows the user to "lock" the top button
and disable the "INST" display while in logging mode. This feature works as follows:
1. The Key Lock option requires enabling from the OdaStat-G™ software package, per the
"Instrument Parameters" section.
TO UNLOCK
First activate LEFT magnetic
switch (LMS) and then press
button while magnet remains
on the switch
2. To Lock the OdaLog® instrument.
Once a logging session is started, the user can select the " LOCK " option from the menu. With
"LOCK" displayed, release the button to lock prompt confirmation. Swipe the magnetic base of the
Allen Tool provided past the right-hand side of the LCD near the "Y" on the screen. The LCD will
display "LOCKED" and the "LOGGING" symbol flashing. The instrument and button operation is
now locked.
3. To Un-Lock the OdaLog® instrument.
Place the magnetic base of the Allen Tool provided on the left-hand side of the LCD display. Press
the top button, and the display will return to the "INST" reading screen. The instrument is now
unlocked.
P/N: 10-1000
.

Figure 13: Accessing Key Lock

Revision G, January 2024
NOTE: ONLY AVAILABLE IF
NOTE: WILL ONLY APPEAR IN
ENABLED IN SOFTWARE AND
MANUAL LOGGING MODE
LOGGING HAS BEEN STARTED
RIGHT MAGNETIC SWITCH (RMS)
To confirm or press button to return
to the INST screen.
TIMEOUT:
IF NO SELECTION IS MADE
AFTER 10 SEC WILL RETURN
TO INST SCREEN
RELEASE
BUTTON
Page | 16