Calibration Errors; Figure 26: Span Calibration; Figure 27: Display Of Failed Calibration - Thermo Scientific Odalog L2 Series User Manual

Table of Contents

Advertisement

OdaLog® L2 – Gas Logger User Manual
3. Ensure that the instantaneous display reads the same as on the cylinder before turning off
the gas regulator and removing the calibration fitting.
4. Ensure that reading returns to close to zero within 5 minutes in clean air and re-zero if
necessary.
Note: The instrument will TIMEOUT and return to the "INST" screen if a SPAN is not done
within 10 minutes.

9.7. Calibration Errors

SENSOR bAd1: The sensor response/reliability has been calculated to be LOW; recheck the
SENSOR bAd2: The sensor response/reliability has been calculated to be VERY LOW; recheck
FAILED:
N/A:
P/N: 10-1000
Figure 23: SPAN Calibration
Figure 24: Display of Failed Calibration
calibration and review Section 7 for sensor information. Span value WILL still
be saved; however, readings may be inaccurate (> ±10 %).
the calibration and review Section 7 for sensor information. Span value WILL
still be saved; however, readings may be inaccurate (> ±10 %).
The sensor has FAILED the response/reliability test; the sensor requires urgent
replacement before being deployed. Span value WILL NOT be saved; return to
the service centre for repair.
If N/A appears on the LCD, there may be a sensor/hardware issue causing
readings to fail. If the error persists, return for service. (ERROR 3 may also
indicate this issue).