10.8
SETTING A
0
Parameter which allows modification of the value
of A
for all cycles except CLOSTRIDIUM.
0
The Set A
menu is accessed using buttons P4 and P6.
0
After accessing SETA
modification mode: the A
number of the program on the first line and the
associated value on the second line.
Scroll with buttons P2 and P8 to change the value.
Confirm with button P5. Press buttons P4 and P6 to
scroll the programs.
The parameter can be edited by the Superuser, from
a minimum of 60 to a maximum of 6000.
Settings are in steps of:
20 for range 60 – 300
300 for range 300 – 6000
The value of A
can be changed for every single
0
program and once changed it remains the same until
changed again (it cannot be automatically reset to
the default value).
10.9

RESET

Reset of device's equipment and configuration.
10.9.1 MC – RESET – CONFIGURATION
Machine configuration.
Parameter which the engineer can force restoration of the
configuration values to the factory values.
This resets all the values except:
-
the relative counters
-
the absolute counters
-
date
-
time
The default setting of the parameter is "UV" (user values)
if the values have been changed from the factory values.
To restore the configuration and equipment values to the
Factory values, set and confirm "FV" (factory values).
[change value with P2/P8, confirm with P5]
User Manual
– Setup A
0
, press P5 to access the A
0
icon (F5) lights up with the
0
193909105 rev.04
0
BPW1260 – BPW4260
Page 68 – 107