Smeg BPW4260 User Manual page 87

Bedpan washers
Table of Contents

Advertisement

ID
DESCRIPTION
AF:73
Internal memory error
Restore fail. Problem on the
AF:78
motherboard.
AF:79
Program not compatible.
Internal memory full. More than
95% of the memory capacity
AF:91
available for storing cycle data
has been used.
Temperature
AF:94
temperature
time. Temperature unstable
Analogue
AF:96
anomaly
Left spray arm (Bedpan 2)
AF:130
anomaly
Right spray arm (Bedpan 1)
AF:131
anomaly
Generator safety thermostat
AF:324
tripped
AF:332
Generator does not fill
User Manual
User: follow the DEFAULT PROCEDURE described above. If
the alarm does not disappear, follow the RESET
PROCEDURE.
User: follow the DEFAULT PROCEDURE described above. If
the alarm does not disappear, follow the RESET
PROCEDURE.
- Pay special attention to the version of the device being
used; e.g. do not activate peristaltic pump P1 when writing
the program if this pump is not installed.
- The program contains parameters not consistent with the
product's configuration. If the program has been
created/customised with the app: check the parameters,
comparing them with those of a Factory program.
Once the limit is reached: do not perform any more cycles
until the memory has been cleared. Contact the after-sales
service.
below
target
User: follow the DEFAULT PROCEDURE described above. If
during
holding
the alarm does not disappear, follow the RESET
PROCEDURE.
Check the water intake and drain connections and that
they are compliant with the installation specifications.
If leaks are noticed around the device: turn off the water
pressure
switch
supply taps.
User: follow the DEFAULT PROCEDURE described above. If
the alarm does not disappear, follow the RESET
PROCEDURE.
Make sure that you have loaded the sanitary aids on the
washing support correctly as instructed to ensure that they
do not obstruct spray arm rotation. Check for any dirt or
limescale in the openings in the spray arm which may be
stopping water from coming out.
Checks and procedures as for AF:130
Make the device safe and contact the after-sales service.
User: follow the DEFAULT PROCEDURE described above. If
the alarm does not disappear, follow the RESET
PROCEDURE.
193909105 rev.04
USER ACTION
BPW1260 – BPW4260
Page 87 – 107

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bpw1260

Table of Contents