Set-Point Entry For Temperature In Two-Door Chambers (Ed, Fd, Fed 720) - Binder 9010-0323 Operating Manual

Table of Contents

Advertisement

6.6.1 Set-point entry for temperature in two-door chambers (ED, FD, FED 720)

With two-door chambers (ED, FD, FED 720) the maximum adjustable temperature set-point depends on
the ambient temperature:
Range from 18 °C up to 26 °C: maximum temperature set-point: 300 °C
Range from above 26 °C up to 40 °C: maximum temperature set-point decreasing with increasing
ambient temperature
Figure 10: Maximum temperature set-point depending on the ambient temperature
Ambient temperature
18 °C up to 26 °C
28 °C
30 °C
32 °C
34 °C
36 °C
38 °C
40 °C
This ensures the maximum lifetime of the
BD / BF / ED / FD / FED (E3.1) 03/2019
Ambient temperature [ °C ]
Maximum temperature set-point
300 °C
298 °C
296 °C
294 °C
292 °C
290 °C
288 °C
286 °C
controller.