Setting The Safety Controller Set-Point - Binder 9010-0323 Operating Manual

Table of Contents

Advertisement

7.3

Setting the safety controller set-point

A limit temperature is entered as the safety controller set-point , i.e. the absolute maximum permitted
temperature value.
Example: Temperature set-point 45 °C, safety controller set-point 50 °C.
Regularly check the safety controller setting relating to the entered temperature set-point
Set the safety controller set-point by approx. 2 °C to 5 °C above the desired temperature set-
point.
4x
without fan
5x
with fan
BD / BF / ED / FD / FED (E3.1) 03/2019
From Normal display
with the arrow-up button to the Safety controller set-point setting menu.
The current safety controller set-point is shown
(class 2 "temperature limiter" or class 3.1 "temperature protection" depending
on the chamber type).
Press OK to enter the safety controller set-point.
The safety controller set-point flashes.
Enter the safety controller set-point with arrow buttons:
10 °C up to 100 °C (with an accuracy of a tenth of a degree) with BD, BF
10 °C up to 300 °C (with an accuracy of one degree) with ED, FD, FED
and confirm with OK.
Back to Normal display.
page 34/106