Figure 2-4: Signal Wiring or Cable
A. Strip wire ends 3/16-in. (4.8 mm) typical
B. Stud size #10
C. Heat shrink tubing 2-in. (50.8 mm) long ½-in. (12.7 mm) size
D. 7.0-in.(177.8 mm) long PTFE tubing, 0.042-in. ID (Cut off drain wire at probe end of
E. ¾ national pipe thread (NPT) hub size, liquid-tight strain relief connector
F. Heat shrink tubing 2-in. (50.8 mm) long ½-in. (12.7 mm) size
G. 8 twisted pairs 24 AWG, stranded, insulated, tinned copper conductors, 400 °F (200 °C),
H. Ferrule, uninsulated
I. Overall cable length by customer 150-ft. (45.7 m) maximum
J. Heat shrink tubing 1-in. (25.4 mm) long, 3/16-in. (4.8 mm) size
K. Typical on both ends of wiring
L. 2.0 ± 0.25 typical
M. For RFI/CE compliance, the connector must provide 360 degrees of electrical contact to
Rosemount OCX8800C
shield.)
300 volts, with overall braid of 34 AWG tinned copper, uninsulated drain wire
the cable shield.
Install
April 2023
33