Emerson Rosemount OCX8800C Manual page 107

Oxygen and combustibles analyzer for hazardous areas
Hide thumbs Also See for Rosemount OCX8800C:
Table of Contents

Advertisement

Manual
00809-0500-4881
Only authorized personnel should use this feature for testing purposes only.
8.4.4 
Configure PlantWeb alert (PWA) simulation from AMS
Procedure
1. Run the Transmitter Options method.
2. From the Transmitter Options menu, select either Simulate Switch or PWA Simulate.
3. Select Enable to enable the simulation feature or Disable to disable the simulation
4. Once the method is complete, select the Simulate PWA tab in the resource block,
5. Now select Simulation on or Simulation off from the Simulate PWA screen.
6. To simulate PlantWeb alerts, use the Simulate PWA screens in the resource block. To
7. Select Device Diagnostics to view the active PWA alarms.
Rosemount OCX8800C
feature.
Figure
8-3.
If the simulation is enabled, the PlantWeb Alarm Simulate parameter is
configurable; otherwise it is read-only.
Figure 8-3: Simulate PWA screen
When Fieldbus simulation is on, the Simulation Switch LED is illuminated. When
PWA simulation is off, the PWA Simulate LED is illuminated.
If PWA simulation is on, all PWA active parameters and resource and transducer
block status parameters are configurable. Otherwise they are read-only.
simulate block alarms, use Simulate TB Temperature Sensor - Temperature Status,
Simulate TB Temperature Status, or Simulate TB Calibration Device - FF Status in the
transducer block.
F
Fieldbus
OUNDATION
April 2023
107