00809-0500-4881
Table 8-30: Input Selector Function Block System Parameters
Parameter name
Index number
SELECTED
21
SELECT_TYPE
19
STATUS_OPTS
10
STRATEGY
3
ST_REV
1
TAG_DESC
2
UPDATE_EVT
23
Figure 8-29: Input Selector Function Block Schematic
A. Selection algorithm
B. Out
C. Selector
8.12.1 
Block errors
Table 8-31
Table 8-31: Block Error Conditions
Condition
number
0
Rosemount OCX8800C
Units
None
None
None
None
None
None
None
lists the conditions reported in the BLOCK_ERR parameter.
Condition name and description
Other: The output has a quality of Uncertain.
F
(continued)
Description
The selected input number (1-4).
Specifies selection method (see Block
Execution).
Allows selection of options for status handling
and processing. The supported status option
for the PID block is Target to Manual if Bad IN.
The Strategy field can be used to identify
grouping of blocks. Theis data is not checked
or processed by the block.
The revision level of the static data associated
with the function block. The revision value will
be incremented each time a static parameter
value in the block is changed.
The user description of the intended application
of the block.
This alert is generated by any change to the
static data.
Fieldbus
OUNDATION
April 2023