Oxygen and combustibles analyzer for hazardous areas
Hide thumbs Also See for Rosemount OCX8800C:
Table of Contents

Advertisement

Maintenance and service
April 2023
4. Use MEK or methylene chloride solvent to clean thread sealant residue from the
Diassemble O
 CAUTION
Equipment damage
Removing the O
Do not remove the O
238
internal pipe threads in the housing.
Refer to the applicable material safety data sheet (MSDS) for solvent handling
precautions.
cell and heater strut assembly
2
cell may damage the cell and platinum pad.
2
cell unless you are certain it needs to be replaced.
2
Manual
00809-0500-4881
Emerson.com/Rosemount