Table 8-14: COe out Tracks Enumerations
COZERO_OUTTRAK description
O
2
CO
2
Both
Table 8-15: Detailed Status
Alarm number
Description
0
No alarm active
1
O
2
O
3
O
4
O
5
O
6
O
7
O
8
O
9
O
10
O
11
Combustibles cell error
12
Combustibles thermocouple open
13
Combustibles thermocouple shorted
14
Combustibles thermocouple reversed
15
Combustibles temperature low
16
Combustibles temperature high
17
Combustibles temperature very high
18
Combustibles heater failure
19
Combustibles heater ramp rate
20
Sample block thermocouple open
21
Sample block thermocouple shorted
22
Sample block thermocouple reversed
23
Sample block temperature low
24
Sample block temperature high
25
Sample block temperature very high
26
Sample block heater failure
27
Sample block heater ramp rate
28
O
29
Combustibles calibration failed
120
cell open
2
cell impedance high
2
thermocouple open
2
thermocouple shorted
2
thermocouple reversed
2
cell temperature low
2
cell temperature high
2
cell temperature very high
2
heater failure
2
heater ramp rate
2
calibration failed
2
(continued)
Value of XD_ERROr (see FF-903)
N/A
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Electronics failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Electronics failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Mechanical failure
Electronics failure
Calibration error
Calibration error