Emerson Rosemount OCX8800C Manual page 88

Oxygen and combustibles analyzer for hazardous areas
Hide thumbs Also See for Rosemount OCX8800C:
Table of Contents

Advertisement

Field Communicator
April 2023
Figure 7-1: Field Communicator Connections - HART
A. Loop connectors
B. Field Communicator
C. Lead set
D. 4-20 mA terminal block in electronics housing
E. O
F. Analog output device
G. Field Communicator rear panel
7.2.2 
Connecting to a F
Connect the Field Communicator with the supplied lead set in parallel with the device to
a Fieldbus segment,
sensitive; an error message displays if the device is connected incorrectly.
 WARNING
EXPLOSION
Explosions can result in death or serious injury.
Do not make connections to the Field Communicator's serial port, digital signal line, or
NiCad recharger jack in an explosive atmosphere.
88
4-20 mA signal line
2
OUNDATION
Figure
7-2. The Field Communicator Fieldbus connections are polarity
Fieldbus segment
Emerson.com/Rosemount
Manual
00809-0500-4881