Emerson Rosemount OCX8800C Manual page 234

Oxygen and combustibles analyzer for hazardous areas
Hide thumbs Also See for Rosemount OCX8800C:
Table of Contents

Advertisement

Maintenance and service
April 2023
5. Loosen clamp screw of COe band heater (M) until the heater rotates freely on the
6. See
7. With one wrench holding the eductor elbow (N,
8. Slide the band heater (M) and heater insulator (K) from the sensor holder (J).
234
sensor holder.
Figure
10-20. Using a straight edge on the sensor holder flat, as shown,
matchmark the upper flange of sensor housing to show the correct alignment of
the sensor holder (J).
Figure 10-20: Alignment of COe Sensor Assembly
A. Matchmark
B. Straightedge
C. Sensor holder flat
flats of sensor holder (J), unthread and remove the COe sensor assembly (I).
Do not allow the eductor elbow to turn.
00809-0500-4881
Figure
10-19) and one wrench on