4. Invert the sensor holder (E,
5. Install and tighten the plug.
6. Lubricate and install the COe sensor gasket (F). Apply anti-seize compound to
7. Install COe sensor holder (A), lockwashers (I), and screws (J). Rotate flat of COe
Rosemount OCX8800C
Figure 10-8: COe Sensor Assembly, Top View
A. Sensor holder flat
B. Sensor holder
C. COe sensor flat
D. Pre-heater flat
E. Sensor holder flat
into plug port. Press down on stainless steel balls and tap sensor holder with plastic
hammer to compact balls in pre-heater chamber.
 CAUTION
Equipment damage
The resistance temperature device (RTD) elements are fragile, and correct alignment
in sensor holder is required for proper Rosemount OCX8800 operation.
Use care when installing the combustibles (COe) sensor.
threads of screws.
sensor (A) to center of sensor holder (E).
Figure
10-24) in a vise and pour stainless steel balls (C)