TROUBLE SHOOTING

4.2.2 Description of the Error Message System

The error message, "error" (M11), is shown in the "time" (A-3) display, (see Figure 39). For more
detailed information, refer to "Table 7: Error Messages", (see Appendix P. VIII).
Figure 39

4.2.3 Procedure while error 14

If Error 14 occurs, there is a problem with the speed sensor. The centrifuge lid is closed for
undefined period of time and in the "rpm/rcf" (A-1) display shows the lettering "USEr GuidE" (see
Figure 40).
Figure 40
To reopen the centrifuge lid, switch off the device and wait until the rotor has come to a standstill.
Take from "Table 6: Acceleration and Deceleration Times" the maximum deceleration time of the
respective rotor. Level 0 corresponds to unbreaked rundown, which occurs at error 14. If the
centrifuge lid is opened before standstill of the rotor, a following error can occur.
Once the rotor has come to a standstill, open the centrifuge lid with the emergency release. Proceed
as described in chapter 4.2.1, p 29. After opening the centrifuge lid, switch on the device again.
Error 14 and the lettering "USEr GuidE" should be eliminated.