6.2.2 Description of the Error Message System

The error message, "error" (M11), is shown in the "time" (A-3) display, (see Figure 25). For more
detailed information, refer to Table 4: , (see Appendix P.V).
Figure 25

6.2.3 Procedure while error 14

If Error 14 occurs, there is a problem with the speed sensor. The centrifuge lid is closed for
undefined period of time and in the "rpm/rcf" (A-1) display shows the lettering "USEr GuidE" (see
FIgure 26).
FIgure 26
To reopen the centrifuge lid, switch off the device and wait until the rotor has come to a standstill!
Take from chapter 4.3.3 the maximum deceleration time of the respective rotor. Level 0 corresponds
to unbreaked rundown, which occurs at error 14. If the centrifuge lid is opened before standstill of
the rotor, a following error can occur.
Once the rotor has come to a standstill, open the centrifuge lid with the emergency release. Proceed
as described in chapter 6.2.1, P. 24. After opening the centrifuge lid, switch on the device again.
Error 14 and the lettering "USEr GuidE" should be eliminated.
If this error occurs again, please contact the service.