4.2.2 Description of the Error Message System

The error message "error" (M11). is shown in the "time" (A-3) display (see 36). For more detailed
information, refer to Table 6: Error Messages (see Appendix P.VIII).
Figure 336

4.2.3 Procedure while error 14 from software version 1.76

If Error 14 occurs, there is a problem with the speed sensor. The centrifuge lid is closed for
undefined period of time and in the "rpm/rcf" (A-1) display shows the lettering "USEr GuidE" (see
Figure 37).
Figure 37
To reopen the centrifuge lid, switch off the device and wait until the rotor has come to a standstill.
Take from "
the respective rotor. Level 0 corresponds to unbreaked rundown, which occurs at error 14. If the
centrifuge lid is opened before standstill of the rotor, a following error can occur.
Once the rotor has come to a standstill, open the centrifuge lid with the emergency release. Proceed
as described in chapter 4.2.1. p 26. After opening the centrifuge lid, switch on the device again.
Error 14 and the lettering "USEr GuidE" should be eliminated.
© Hermle Labortechnik GmbH
Table 5: Acceleration and Deceleration Times
ZK496_V1.22_eng