MAINTENANCE
ATTENTION: Before applying any other, cleaning resp. decontamination method, other than
what was recommended by the manufacturer, contact the manufacturer to ensure that it will
not damage the unit or the rotor.

3.1.6 Glass Breakage

With high g-values, the rate of glass tube breakage increases. Glass splinters have to be removed
immediately from rotor, buckets, adapters and the rotor chamber itself. Fine glass splinters will
scratch and therefore damage the protective surface coating of a rotor. If glass splinters remain in
the rotor chamber, fine metal dust will build up, due to air circulation. This very fine, black metal dust
will severely pollute the rotor chamber, the rotor, the buckets, and the samples.
If necessary, replace the adapters, tubes and accessories, to avoid further damage. Check the rotor
bores regularly, for residue and damage.
ATTENTION: Please check the relevant specifications of the tubes with the manufacturer!

3.2 Lifetime of Rotors, Round and Rectangular Buckets, Accessories

Rotors and rotor lid made of aluminum or stainless steel, have a operating time of max. 7 years
from first use.
Transparent rotor lids and caps, made of PC or PP, as well as rotors, tube racks and adapters of
PP, have a maximum operating time of up to 3 years, from first time use.
Conditions for the Operating Time:
Proper use, damage-free condition, recommended care.
For additional information on aerosol-tight rotors, lids and buckets, please see below
The aerosol tightness of rotors, rotor lids, buckets and caps have been tested and certified by the "TÜV Nord
CERT GmbH, Certification Body Consumer Products, Essen (Germany)", in accordance with Annex AA of IEC
61010-2-020. The certificates can be downloaded on our webpage, www.hermle-labortechnik.de. Aerosol-
tight rotors and buckets are marked with the label, "aerosol-tight".
ATTENTION: Autoclaving, mechanical stresses and contamination, by chemicals or other
aggressive solvents, can impair the aerosol-tightness of the rotors and buckets.
-
Check the integrity of the seals of the aerosol-tight rotor lids or caps, before each use.
-
Use only aerosol-tight rotor lids or caps, if the seals are undamaged and clean.
-
Aerosol-tight plastic rotors as well as plastic lids and seals of aerosol-tight aluminum rotors and
buckets must be replaced after 10 autoclaving cycles.
-
Never store aerosol-tight rotors or buckets closed.
-
Plastic lids and seals of aerosol-tight aluminum rotors and aluminum buckets must be replaced after ten
autoclaving cycles.
28
Z327K_V1.22_eng
© Hermle Labortechnik GmbH